Chapter 5
Using the SCXI-1141/1142/1143 Module
© National Instruments Corporation
5-13
Calibration
The SCXI-1141/1142/1143 is shipped with a calibration certificate and is 
calibrated at the factory to the specifications described in Appendix A, 
. Calibration constants are stored inside the calibration 
EEPROM and provide software correction values your application 
development software uses to correct the measurements for both offset and 
gain errors in the module.
External Calibration
If you have an accurate calibrator and DMM, you can externally calibrate 
the SCXI-1141/1142/1143 gain and offset constants using NI-DAQmx 
functions.
Most external calibration documents for SCXI modules are available to 
download from 
ni.com/calibration
 by clicking Manual Calibration 
Procedures. For external calibration of modules not listed there, Basic 
Calibration Service or Detailed Calibration Service is recommended. 
You can get information about both of these calibration services from 
ni.com/calibration
. NI recommends performing an external 
calibration once a year.