Chapter 1
Taking Measurements with the NI 4050
© National Instruments Corporation
1-7
Figure 1-5.  Connections for Resistance Measurement
2.
Select 2-wire resistance mode.
3.
Select the range for your measurement—200 
Ω
, 2 k
Ω
, 20 k
Ω
, 200 k
Ω
2 M
Ω
, 200 M
Ω
, or autorange.
The value indicator indicates the resistance measured. See the 
, fo
more information on 2-wire resistance measurements.
Measuring the Voltage Drop Across a Diode
The NI 4050 can excite a device under test and read the resulting voltage 
drop. Diode mode is useful for testing diodes. Use the following procedure 
to measure the forward drop across a diode. Voltage up to 2 V can be 
measured in this mode.
1.
Connect the test probes to a diode as shown in Figure 1-6. To 
accurately measure the forward voltage of a diode, make sure that the 
diode is not connected to any other circuits. The NI 4050 biases the 
diode with a current of 100
µ
A and measures the resulting voltage 
drop. Diode measurements are made with a fixed range of 2.0 V.
Figure 1-6.  Connecting Signals for Diode Test
Resistor
250 V
MAX.
LO
HI
Diode
+
100 
µ
A
250 V
MAX.
LO