Tool Probe Calibration Cycle
This is used to set the Z datum for length preset, the effective probe 
stylus diameter for setting tool diameter registers, and establishes 
the center of the probe stylus. 
This cycle is used to calibrate the probe. This is used to set the Z 
datum for length preset, establishing the center of the probe stylus, 
and the effective probe stylus diameter for setting tool diameter 
registers.
Calibration must be done at least once before using the 
tool probe. Once the probe has been calibrated, calibration 
does not need to be done again unless the probe is moved 
or a new part is being setup. The cycle must always know 
the relationship between the top of the part and the top of 
the probe to set the TLO. 
Field
Code
Description
DiamOfStd
D
The diameter of the part of the calibration 
standard that comes in contact with the 
probe stylus during calibration. This should 
be an exact measurement. (Optional) 
override for the 
DiameterOfToolProbeGauge machine 
setup parameter) .
DistDown
E
The incremental distance from the current Z 
Retract amount to go down along the side 
of the probe stylus when doing a diameter 
pick. The maximum E value is 0.55" (13.97 
mm) or the tool may crash into the probe or 
table. If you enter a value larger than 0.55" 
(13.97 mm), the control issues an error 
message. If E is not set, the cycle uses a 
default value of 0.1" (2.54 mm). (Optional) 
[Default: 0.1"]. Z Retract Amount is set in 
the Tool Probe Parameters.