NI PXI/PCI-5114 Specifications
14
ni.com
Calibration
Power
Specification
Value
Self-Calibration
Self-calibration is done on software command. The calibration
corrects for gain, offset, compensated 1 M
corrects for gain, offset, compensated 1 M
Ω attenuator, triggering,
and timing adjustment errors for all input ranges.
External Calibration
(Factory Calibration)
(Factory Calibration)
The external calibration calibrates the VCXO, gain, and the voltage
reference. Appropriate constants are stored in nonvolatile memory.
reference. Appropriate constants are stored in nonvolatile memory.
Interval for External
Calibration
Calibration
2 years
Warm-Up Time
15 minutes
Specification
Typical Value
+3.3 VDC
NI PXI-5114
NI PCI-5114
840 mA
1.6 A
+5 VDC
1.1 A
1.7 A
+12 VDC
250 mA
45 mA
–12 VDC
170 mA
—
Total Power
13.32 W
14.32 W