National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册

下载
页码 68
© National Instruments Corporation
15
NI 4070/4072 Calibration Procedure
71. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_VOLTS
Range = 
300
Resolution = 
300e–6
72. Call 
niDMM_Read
. Before you apply the voltage, the DMM must be in 
the 300 V range.
73. Output 300 V on the calibrator.
74. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 15.
75. Output –300 V on the calibrator.
76. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 15.
77. Reset the calibrator for safety reasons.
You have completed verifying the DC voltage of the NI 4070/4072. Select 
one of the following options:
If you want to continue verifying other modes, go to the Verifying AC 
Voltage
 sec
tion.
If you do not want to verify other modes and you are performing a 
post-adjustment verification, go to the 
If you do not want to verify any additional modes and you are 
performing a pre-adjustment verification, call 
niDMM_close
 to close 
the session.
Verifying AC Voltage 
To verify AC voltage of the NI 4070/4072, complete the following steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of the Fluke 5440 cable to the 
appropriate banana plug connectors on the NI 4070/4072, and fasten 
the connectors on the other end of the cable to the appropriate 
calibrator binding posts. Figure 1 shows the correct connections. 
Table lists the cable connections.
3.
Output 5 mV at 1 kHz on the calibrator.
4.
Call 
niDMM_reset
 to reset the NI 4070/4072 to a known state.