National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册
© National Instruments Corporation
31
NI 4070/4072 Calibration Procedure
3.
Call
niDMM_reset
to reset the NI 4070/4072 to a known state.
4.
Set the current output on the calibrator to NORM and output 0 A.
5.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_DC_CURRENT
•
Range =
0.02
•
Resolution =
20e–9
6.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 19.
7.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_DC_CURRENT
•
Range =
0.2
•
Resolution =
200e–9
8.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 19.
9.
Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_DC_CURRENT
•
Range =
1
•
Resolution =
1e–6
10. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 19.
11. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_DC_CURRENT
•
Range =
0.02
•
Resolution =
20e–9
12. Call
niDMM_Read
to configure the NI 4070/4072 for a current mode
before applying current.
Table 9. Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
(Fluke 5440 Cable)
Binding Post
(Fluke 5700A/5720A Calibrator)
HI SENSE
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD