NI 4070/4072 Calibration Procedure
3.
Call 
niDMM_reset
 to reset the NI 4070/4072 to a known state.
4.
Set the current output on the calibrator to NORM and output 0 A.
5.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_CURRENT
Range = 
0.02
Resolution = 
20e–9
6.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 19.
7.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_CURRENT
Range = 
0.2
Resolution = 
200e–9
8.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 19.
9.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_CURRENT
Range = 
1
Resolution = 
1e–6
10. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 19.
11. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_DC_CURRENT
Range = 
0.02
Resolution = 
20e–9
12. Call 
niDMM_Read
 to configure the NI 4070/4072 for a current mode 
before applying current.
Table 9.  Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post
(Fluke 5700A/5720A Calibrator)
HI SENSE
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD