NI 4070/4072 Calibration Procedure
34
ni.com
14. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_AC_CURRENT
Range = 
1
Resolution = 
1e–6
15. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 20.
16. Output 1 A at 1 kHz on the calibrator.
17. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 20.
You have completed verifying the AC current of the NI 4070/4072. Select 
one of the following options:
If you want to continue verifying other modes, go to thVerifying 
Frequency
 section
.
If you do not want to verify other modes and you are performing a 
post-adjustment verification, go to the 
 section. 
If you do not want to verify any additional modes and you are 
performing a pre-adjustment verification, call 
niDMM_close
 to close 
the session.
Verifying Frequency
Notes
The frequency of the NI 4070/4072 is not user adjustable. If this verification 
procedure indicates that the frequency is out of specification, return the NI 4070/4072 
to NI for repair.
This verification procedure is optional and requires additional test 
equipment. If you do not want to verify frequency, select one of the 
following options:
If you are calibrating an NI 4072 and want to continue verifying other 
modes, go to the 
If you do not want to verify other modes and are performing a 
post-adjustment verification, go to the 
 section.
If you do not want to verify any additional modes and you are 
performing a pre-adjustment verification, call 
niDMM_close
 to close 
the session.