NI 4070/4072 Calibration Procedure
Verifying Capacitance and Inductance (NI 4072 Only)
This verification procedure only applies to the NI 4072 and requires 
additional test equipment, as indicated in the 
Note
The NI 4072 inductance accuracy is theoretically verified if the capacitance 
accuracy meets the specifications. If you have access to precision inductors, you can verify 
the inductance measurements by comparing your results with the published accuracy 
specifications.
NI suggests using traceable capacitor standards with low thermal drift. You 
can use different verification capacitors to verify each capacitance range. 
You can verify two ranges with the same verification capacitor as long as 
its value is 
≥10% of the higher capacitor range. For example, you can use a 
1 nF verification capacitor to test both the 10 nF and 1 nF ranges.
After taking each measurement, verify that the measurement falls between 
the tolerances listed in Table 22. Tolerances are provided instead of 
absolute limits, because you can use capacitance verification values other 
than the values suggested, or the calibrated value may differ slightly from 
the nominal capacitance (for example, 272.43 pF instead of 270.00 pF). 
The tolerances shown in Table 22 correspond to the NI 4072 accuracy 
specifications.
The following verification procedure assumes the use of verification 
capacitors with the following values: 270 pF, 1 nF, 100 nF, 10
μF, and 
1000
μF.
The configuration of the cables and fixtures should be consistent 
throughout each measurement. If you are using cables to connect the 
verification capacitors to the NI 4072 banana plug connectors, minimize 
noise by ensuring that the cables remain fixed and do not move during the 
measurement. 
Keep direct contact with the verification capacitors to a minimum so that 
they are constantly kept at the ambient temperature. After connecting a 
capacitor to the NI 4072 terminals, NI recommends waiting 30 seconds for 
the capacitor temperature to stabilize.
Note
You should know the total capacitance up to the banana connectors that plug into the 
NI 4072 before performing the verification procedure.