National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册
© National Instruments Corporation
53
NI 4070/4072 Calibration Procedure
9.
Reset the calibrator for safety reasons.
10. Call
niDMM_CalAdjustMisc
with Type set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the AC voltage modes of the NI 4070/4072.
Select one of the following options:
Select one of the following options:
•
If you are performing additional adjustments, refer to the following
sections, as applicable:
sections, as applicable:
–
–
Caution
For the NI 4072, adjusting the capacitance and inductance is required. Skipping
this step causes an incorrect adjustment of the device.
•
If you are not performing additional adjustments, refer to one of the
following sections:
following sections:
–
—to verify your new calibration
coefficients before saving them to the EEPROM
–
—if you do not want to
verify the adjustments you have just made
Adjusting Current Modes
If you do not use the current modes (DC and AC), or the accuracy is
insignificant for your application, you can skip this section and select one
of the following options:
insignificant for your application, you can skip this section and select one
of the following options:
•
If you skip this section and you are calibrating an NI 4072, go to the
section.
38
100 V
200
NIDMM_VAL_AC_VOLTS
300
200e3
100 V
200
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
200e3
39
50 V
300
NIDMM_VAL_AC_VOLTS
300
300e3
50 V
300
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
300e3
40
10 V
500
NIDMM_VAL_AC_VOLTS
300
500e3
10 V
500
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
500e3
Table 13. niDMM_CalAdjustACFilter Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustACFilter Parameters
Amplitude
Frequency
(kHz)
Mode
Range
(V)
Frequency
(Hz)