NI 4070/4072 Calibration Procedure
9.
Reset the calibrator for safety reasons.
10. Call 
niDMM_CalAdjustMisc
 with Type set to 
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the AC voltage modes of the NI 4070/4072. 
Select one of the following options:
If you are performing additional adjustments, refer to the following 
sections, as applicable:
Caution
For the NI 4072, adjusting the capacitance and inductance is required. Skipping 
this step causes an incorrect adjustment of the device.
If you are not performing additional adjustments, refer to one of the 
following sections:
to verify your new calibration 
coefficients before saving them to the EEPROM
if you do not want to 
verify the adjustments you have just made
Adjusting Current Modes
If you do not use the current modes (DC and AC), or the accuracy is 
insignificant for your application, you can skip this section and select one 
of the following options:
If you skip this section and you are calibrating an NI 4072, go to the 
 section.
38
100 V 
200
NIDMM_VAL_AC_VOLTS
300
200e3
100 V 
200
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
200e3
39
50 V 
300
NIDMM_VAL_AC_VOLTS
300
300e3
50 V 
300
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
300e3
40
10 V
500
NIDMM_VAL_AC_VOLTS
300
500e3
10 V
500
NIDMM_VAL_AC_VOLTS_DCCOUPLED
300
500e3
Table 13.  niDMM_CalAdjustACFilter Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustACFilter Parameters
Amplitude
Frequency 
(kHz)
Mode
Range 
(V)
Frequency 
(Hz)