NI 4070/4072 Calibration Procedure
58
ni.com
To finish the calibration and close the session, go to the Completing the 
Adjustment Procedures 
section.
Completing the Adjustment Procedures
To complete the adjustment procedure for the NI 4070/4072 and close the 
session, call 
niDMM_CloseExtCal
 with the following parameter:
Action = 
NIDMM_EXTCAL_ACTION_SAVE
 if the results of the 
calibration were satisfactory and you want to save the new calibration 
coefficients to the EEPROM.
Otherwise,
Action = 
NIDMM_EXTCAL_ACTION_ABORT
 if the results of the 
calibration were unsatisfactory and you want to restore the original 
calibration coefficients to the EEPROM.
Verification Limits
This section includes the verification limits for DC voltage, AC voltage, 
4-wire resistance, 2-wire resistance, DC current, AC current, and frequency 
for the NI 4070/4072and the verification tolerances for capacitance on the 
NI 4072. Compare these limits to the results you obtain in the 
 section.
Note
Use the values in the 24-Hour Limits column for a post-adjustment verification only. 
Otherwise, use the values in the 2-Year Limits column.
Limits in the following tables are based upon the February 2007 edition of the 
NI 4070/4072 Specifications. Refer to the most recent NI 4070/4072 specifications online 
at 
ni.com/manuals
. If a more recent edition of the specifications is available, recalculate 
the limits based upon the latest specifications.
DC Voltage
Table 15.  NI 4070/4072 DC Voltage Verification Limits
Calibrator 
Amplitude
Range
Input Resistance
2-Year Limits
24-Hour Limits
Lower
Upper
Lower
Upper
0 V
1 V
>10 G
Ω/10 MΩ
–6
μV
6
μV
–2
μV
2
μV
0 V
10 V
>10 G
Ω
/
10 M
Ω
–60
μV
60
μV
–20
μV
20
μV
0 V
100 V
10 M
Ω
–600
μV
600
μV
–200
μV
200
μV
0 V
300 V
10 M
Ω
–6 mV
6 mV
–1.8 mV
1.8 mV
100 mV
100 mV
>10 G
Ω
/
10 M
Ω
0.099994 V
0.100006 V
0.099998 V
0.100002 V