National Instruments NI 4070/4072 6 1/2-Digit FlexDMM 用户手册

下载
页码 68
© National Instruments Corporation
61
NI 4070/4072 Calibration Procedure
4-Wire Resistance
Note
Tolerances are provided for 4-wire resistance instead of absolute limits because the 
limits depend on the actual resistance value output by your calibrator.
2-Wire Resistance
Note
Tolerances are provided for 2-wire resistance instead of absolute limits because the 
limits depend on the actual resistance value output by your calibrator.
Table 17.  NI 4070/4072 4-Wire Resistance Verification Tolerances
Calibrator
Resistance
Range
2-Year Tolerance
(ppm of Range)
24-Hour Tolerance 
(ppm of Range)
10 M
Ω
10 M
Ω
±410 ppm
±102 ppm
1 M
Ω
1 M
Ω
±100 ppm
±22 ppm
100 k
Ω
100 k
Ω
±86 ppm
±17 ppm
10 k
Ω
10 k
Ω
±83 ppm
±14 ppm
1 k
Ω
1 k
Ω
±83 ppm
±14 ppm
100
Ω
100
Ω
±90 ppm
±25 ppm
0
Ω
10 M
Ω
±10 ppm
±2 ppm
0
Ω
1 M
Ω
±10 ppm
±2 ppm
0
Ω
100 k
Ω
±6 ppm
±2 ppm
0
Ω
10 k
Ω
±3 ppm
±2 ppm
0
Ω
1 k
Ω
±3 ppm
±2 ppm
0
Ω
100
Ω
±10 ppm
±10 ppm
Table 18.  NI 4070/4072 2-Wire Resistance Verification Tolerances
Calibrator
Resistance
Range
2-Year Tolerance
(ppm of Range)
24-Hour Tolerance 
(ppm of Range)
0
Ω
10 k
Ω
±40 ppm
±20 ppm
0
Ω
1 k
Ω
±400 ppm
±200 ppm
0
Ω
100
Ω
±4000 ppm
±2000 ppm
100 M
Ω
100 M
Ω
±6040 ppm
±920 ppm
10 M
Ω
10 M
Ω
±410 ppm
±102 ppm
1 M
Ω
1 M
Ω
±100 ppm
±22 ppm