NI 4070/4072 Calibration Procedure
64
ni.com
Appendix A: Calibration Options
The complete calibration process for the NI 4070/4072 consists of 
verifying, adjusting, and reverifying a device. During verification, you 
compare the measured performance to an external standard of known 
measurement uncertainty to confirm that the product meets or exceeds 
specifications. Figure 4 shows the procedural flow for verification.
During adjustment, you correct the measurement error of the device by 
adjusting the calibration constants and storing the new calibration constants 
in the EEPROM. Frequency is the only mode that does not require 
adjustment. Figure 5 shows the procedural flow for adjustment. 
Reverifying all modes after adjustments ensures that the adjustment 
procedures were performed correctly.
Normally, the calibration sequence is as follows:
1.
Verify the operation of the NI 4070/4072 using the 2-year accuracy 
limits (or the 90-day accuracy limits if it has been externally calibrated 
within that time).
2.
Adjust the NI 4070/4072.
3.
Reverify the NI 4070/4072 using the 24-hour accuracy limits (or the 
2-year accuracy limits when the 24-hour limits are not specified).
Depending on your measurement and accuracy requirements, a complete 
calibration of the NI 4070/4072 may not be necessary. A number of options 
are available that can shorten the calibration time. The following 
adjustment options are available:
Complete calibration—Performing the entire calibration procedure 
from beginning to end; guarantees that the NI 4070/4072 performs at 
or above the published specifications for all modes and ranges
Complete calibration with exceptions:
Omitting AC voltage mode steps if you do not use the AC voltage 
modes or if the AC voltage accuracy is irrelevant
Omitting DC/AC current mode steps if you do not use the current 
modes or if the DC/AC current accuracy is irrelevant
Omitting both AC voltage and DC/AC current mode steps if you 
do not use those modes or if the accuracy of those measurements 
is irrelevant