VXI-MXI-2 Configuration and Installation
3-14
ni.com
Use switch 2 of U35 to select whether you want the VXI-MXI-2 to 
automatically control termination of the MXIbus. Switch 1 of U35 lets 
you manually control whether to terminate the MXIbus when automatic 
termination is turned off. Switch 1 has no effect when switch 2 is set for 
automatic MXIbus termination; you must turn off automatic termination 
if you want to manually control termination.
Figure 3-9.  MXIbus Termination
Configuration EEPROM
The VXI-MXI-2 has an onboard EEPROM, which stores default register 
values that are loaded at power-on. The EEPROM is divided into two 
halves—a factory-configuration half, and a user-configuration half. Both 
halves were factory configured with the same configuration values so you 
can modify the user-configurable half, while the factory-configured half 
stores a back-up of the default user settings.