6-14  Calibration
NetScan User’s Manual
2.
 
Apply calibration voltage from the precision calibrator to the first channel of each card (see following
table), and verify with a digital multimeter.  Requirements for both devices are specified in Required
Equipment
.
Volts DC
Slot # 1
Slot #  2
Slot #  3
Slot #  4
Slot # 5
Slot #  6
Slot #  7
Slot # 8
-50 mV
G1,1X
G17,1X
G33,1X
G49,1X
G65,1X
G81,1X
G97,1X
G113,1X
+50 mV
G1,1X
G17,1X
G33,1X
G49,1X
G65,1X
G81,1X
G97,1X
G113,1X
-100 mV
G1,11X
G17,11X
G33,11X
G49,11X
G65,11X
G81,11X
G97,11X
G113,11X
+100 mV
G1,11X
G17,11X
G33,11X
G49,11X
G65,11X
G81,11X
G97,11X
G113,11X
-1.0 mV
G1,12X
G17,12X
G33,12X
G49,12X
G65,12X
G81,12X
G97,12X
G113,12X
+1.0 mV
G1,12X
G17,12X
G33,12X
G49,12X
G65,12X
G81,12X
G97,12X
G113,12X
-5.0 mV
G1,13X
G17,13X
G33,13X
G49,13X
G65,13X
G81,13X
G97,13X
G113,13X
+5.0 mV
G1,13X
G17,13X
G33,13X
G49,13X
G65,13X
G81,13X
G97,13X
G113,13X
-10 mV
G1,14X
G17,14X
G33,14X
G49,14X
G65,14X
G81,14X
G97,14X
G113,14X
+10 mV
G1,14X
G17,14X
G33,14X
G49,14X
G65,14X
G81,14X
G97,14X
G113,14X
Note:  In the slot columns, the digit after the G is the channel number and the number immediately preceding
the X assigns the voltage value, e.g., G1,1X means “Calibrate Gain for channel 1, ±50 mV.”
3.
 
After each calibration type 
U2X
 and check for a return of  
U128
 to confirm no errors.
Response codes and meanings applicable to calibration are as follows:
Response to 
U2X
Indication for Calibration Mode
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature sensor out of range)
U144
Invalid password
U160
Invalid command
 
4.
 
After completing the gain calibration of all cards, type 
EX
 to end the calibration mode.
5.
 
Type 
U2X
 to confirm calibration mode is disabled.  A return of 
U000
 confirms this.
 
Note: You can view the calibration constants by typing an entry similar to the following:
 
C#1X
  
QC?    
…for slot #1
 
You can monitor the calibration voltage by typing an entry similar to the following:
 
*B        
to clear the buffer
 
C1,1X     
to addresses the channel, this example is for slot #1, channel 1
 
T1,1,0,0X;@X 
… to trigger the acquisition (Trigger LED flashes); ); 
@X
 starts acquisition
 
U13X 
every entry of U13X results in an updated value of voltage in counts
 
Note that 
@X 
can be used to start or stop the acquisition.