Emerson Network Router 用户手册

下载
页码 110
62
ProLink
®
 II Software for Micro Motion
®
 Transmitters
Transmitter Configuration, Characterization, and Calibration
Figure 5-6
D3 or D3 and D4 density calibration
Fill sensor with D3 fluid
Calibration
Density cal - Point 3
Do Cal
Enter density of D3 fluid
Fill sensor with D4 fluid
Calibration in Progress 
light turns red
ProLink
Calibration in Progress 
light turns green
Calibration
Density cal - Point 4
Do Cal
Enter density of D4 fluid
Calibration in Progress 
light turns red
ProLink
Calibration in Progress 
light turns green
Close shutoff valve 
downstream from sensor
Done
Close
Close