Adjusting and
Maintaining the Printer
5.0 Media Sensor Calibration
 In addition to the Sensor Type selection, Media Sensor calibration ensures that
each label is detected correctly and reliably. Perform calibration when  the
‘Uncalibrated’ message is displayed. Two different methods are available to
calibrate the printer: Standard and Advanced Entry.
 
 
5.0.1
Standard Calibration
Standard Calibration is appropriate for most media types. During the process,
the printhead assembly can be raised for visual access to the media and media
sensor. In addition, displayed sensor readings can be used to indicate the best
position over the media, which is helpful when using small, position-critical
TOF notches or marks. Three readings are required:
  Empty:  
No media in the sensor.
 
  Gap (or Mark): Only the backing, notch, or reflective mark in the sensor.
 
  Paper: 
The label (with the liner attached) in the sensor.
 
 To perform a Standard Calibration:
 Step
 Operator Action
 Displayed Message
 Comment
 
 1
 
 
 (Media should be loaded
and the appropriate
Sensor Type selected.)
 
 Turn ‘On’ the printer.
 
 
 
 
 
 
 Wait until the
printer initializes
(about six
seconds).
 
 
 2
 
 
 Press the 
 
Key.
 
 Raise the printhead
assembly.
 
 
 
 
 
  
 
 You are entering
Menu Mode.