Chapter 2
Configuring and Installing the SCXI Chassis
© National Instruments Corporation
2-9
Figure 2-7.  Address Setting Examples
SCXI-1000DC
Unless you are using multiple chassis and need to configure one or more 
SCXI chassis for a different address, you can skip this section, and the 
SCXI chassis retains the factory-default address of 0.
You can configure the SCXI chassis for one of 32 different addresses so that 
you can connect multiple SCXI chassis to the same control source. Three 
jumpers that determine the chassis address are located behind the front 
panel of Slot 0 just below the Reset button. The chassis address is the sum 
of the values of all the jumpers. Figure 2-8 shows examples of both the 
factory-default setting of address 0 and the jumper settings for address 19.
Note
SCXI-1000DC chassis through revision C do not have address jumpers or switches 
and respond to any address, but you cannot use them in multichassis systems. Revision D 
and later chassis use jumpers on Slot 0 for chassis addressing.
ON
1
2
3
4
5
Chassis Default
Address = 0
ON
1
2
3
4
5
Address = 19