Chapter 22 
Tips and Maintenance 
Introduction 
This chapter covers basic maintenance procedures that can be performed by the user. For 
complete service, disassembly, repair, and calibration information, see the Service 
Manual. You will find the part number of the Service Manual in section ‘Parts and 
Accessories’ in this chapter. 
Cleaning the Analyzer and its Accessories 
Clean the Analyzer and accessories with a damp cloth and a mild soap. Do not use 
abrasives, solvents, or alcohol. These may damage the text. 
Additional to this it is recommended to open the jaws of the Current Clamp and to wipe 
the magnetic pole pieces with a lightly oiled cloth. This in order to avoid rust or corrosion 
to form on the magnetic poles. 
Storing the Analyzer 
Before storing the Analyzer for an extended period of time, fully charge the NiMH 
battery. 
Keeping the Battery in Good Condition 
When the Analyzer is powered by the battery, the battery condition symbol in the screen 
header informs you about the charge condition. This symbol ranges from fully charged to 
empty: 
 
To keep the battery in optimal condition, you must let it discharge fully and then charge 
it. A full charge takes 4 hours with the Analyzer turned off. Repeat this at least twice a 
year. 
Installation of Options in Fluke 434  
The Advanced Functions Mains Signaling and Logging that are available in Fluke 435, 
can be activated in an existing Fluke 434. Activation can be done by the user done via a 
pin-code that is unique for the serial number of your Analyzer. The code is supplied by 
Fluke. Contact your Fluke sales representative for details on how to obtain your pin-code. 
Extra memory such as present in Fluke 435 can not be added in this way.