selection for either rising or falling edge. The selected edge of TRIG2 
initiates the posttriggered phase of a pretriggered DAQ sequence. In 
pretriggered mode, the TRIG1 signal initiates the acquisition. The scan 
counter indicates the minimum number of scans before TRIG2 can be 
recognized. After the scan counter decrements to zero, it is loaded with the 
number of posttrigger scans to acquire while the acquisition continues. 
The device ignores TRIG2 if it is asserted prior to the scan counter 
decrementing to zero. After the selected edge of TRIG2 is received, the 
device acquires a fixed number of scans and the acquisition stops. This 
mode acquires data both before and after receiving TRIG2.
As an output, TRIG2 reflects the posttrigger in a pretriggered DAQ 
sequence even if another PFI is externally triggering the acquisition. 
TRIG2 is not used in posttriggered data acquisition. The output is an active 
high pulse with a pulse width of 25 to 50 ns. This output is set to 
high-impedance at startup.
Figures 4-16 and 4-17 show the timing requirements for TRIG2.
 
Figure 4-16.  TRIG2 Input Signal Timing
 
Figure 4-17.  TRIG2 Output Signal Timing
Rising-Edge
Polarity
Falling-Edge
Polarity
t
w
 = 10 ns minimum
t
w
t
w
t
= 25 – 50 ns