Chapter 4
Connecting Signals
4-26
ni.com
CONVERT* Signal
Any PFI pin can receive as an input the CONVERT* signal, which is 
available as an output on the PFI2/CONVERT* pin.
Refer to Figures 4-23 and 4-24 for the relationship of CONVERT* to the 
DAQ sequence.
As an input, CONVERT* is configured in the edge-detection mode. You 
can select any PFI pin as the source for CONVERT* and configure the 
polarity selection for either rising or falling edge. The selected edge of 
CONVERT* initiates an A/D conversion.
As an output, CONVERT* reflects the actual convert pulse that is 
connected to the ADC even if another PFI is externally generating the 
conversions. The output is an active low pulse with a pulse width of 
50 to 100 ns. This output is set to high-impedance at startup.
Figures 4-20 and 4-21 show the input and output timing requirements for 
CONVERT*.
 
Figure 4-20.  CONVERT* Input Signal Timing
 
Figure 4-21.  CONVERT* Output Signal Timing
Rising-Edge
Polarity
Falling-Edge
Polarity
t
w
 = 10 ns minimum
t
w
t
w
t
= 50 – 100 ns