National Instruments 1128 用户手册
Chapter 1
Installing and Configuring the SCXI-1127/1128
1-12
ni.com
2.
Connect the HV8-HV8 cable from the first chassis, normally the
connector behind slot 1, to the HVAB connector (behind slot 4) of the
next chassis.
connector behind slot 1, to the HVAB connector (behind slot 4) of the
next chassis.
3.
Connect the SH9MD-9MD cable from the AUX OUT connector of the
first chassis to the AUX IN connector of the next chassis.
first chassis to the AUX IN connector of the next chassis.
4.
Repeat steps 1 through 3 for each additional chassis.
Notes
An SCXI-1127/1128 is required in slot 4 to establish communications with the
chassis. If slot 4 is empty, the system will not operate. It is this module that you must
specify in MAX as the module cabled to the DMM.
specify in MAX as the module cabled to the DMM.
If a chassis is configured with less than four slots of the HVAB-backplane adapter, use the
right-most available HVAB connector to extend to the next chassis. In this instance,
right-most means the slot closest to slot 1 when the chassis is viewed from the back.
right-most available HVAB connector to extend to the next chassis. In this instance,
right-most means the slot closest to slot 1 when the chassis is viewed from the back.
4-Slot and 12-Slot Multichassis Configuration
In addition to the items needed for a single-chassis system, you need the
following items to install this configuration:
following items to install this configuration:
•
SCXI-1358 multichassis expansion kit
–
HV8-HV8 cable
–
SH9MD-9MD cable
–
Three 8-position HVAB plugs
–
One 8-slot HVAB-backplane adapter
–
One 2-slot HVAB-backplane adapter
–
Two 1-slot HVAB-backplane adapters