National Instruments 1128 用户手册

下载
页码 97
Chapter 1
Installing and Configuring the SCXI-1127/1128
1-12
ni.com
2.
Connect the HV8-HV8 cable from the first chassis, normally the 
connector behind slot 1, to the HVAB connector (behind slot 4) of the 
next chassis.
3.
Connect the SH9MD-9MD cable from the AUX OUT connector of the 
first chassis to the AUX IN connector of the next chassis.
4.
Repeat steps 1 through 3 for each additional chassis.
Notes
An SCXI-1127/1128 is required in slot 4 to establish communications with the 
chassis. If slot 4 is empty, the system will not operate. It is this module that you must 
specify in MAX as the module cabled to the DMM.
If a chassis is configured with less than four slots of the HVAB-backplane adapter, use the 
right-most available HVAB connector to extend to the next chassis. In this instance, 
right-most means the slot closest to slot 1 when the chassis is viewed from the back.
4-Slot and 12-Slot Multichassis Configuration
In addition to the items needed for a single-chassis system, you need the 
following items to install this configuration:
SCXI-1358 multichassis expansion kit
HV8-HV8 cable
SH9MD-9MD cable
Three 8-position HVAB plugs
One 8-slot HVAB-backplane adapter
One 2-slot HVAB-backplane adapter
Two 1-slot HVAB-backplane adapters