Vascular
163
 
Hints in General
• Topogram: TOP, 512/1024 or LAT 256
• Patient positioning: 
 
Patient lying in supine position, arms positioned 
comfortably above the head in the head-arm rest, 
lower legs supported.
• Patient respiratory instructions: inspiration. 
• Oral administration of contrast medium:
 
The use of water will not obscure the blood vessels, 
thus allowing CTA post-processing to be performed 
easily afterwards.
• Be careful when examining pheochromocytoma 
patients. Administration of an IV CM injection in 
such cases may trigger a hypertensive crisis!
• To further optimize MPR image quality, we recom-
mend that you reduce one or more of the following 
parameters: collimation, reconstruction increment 
and slice width for image reconstruction.
Head Kernels
• For soft tissue head studies, the standard kernel is 
H41s; softer images are obtained with H31s or H21s, 
sharper images with H50s.
Body Kernels
• As standard kernels for body tissue studies B31s or 
B41s is recommended; softer images are obtained 
with B20s.