Siemens Software version syngo CT 2005C 用户手册
196
Children
HeadSeq_Baby
Indications:
Sequential mode for routine head studies for children,
e.g., tumors, hydrocephalus, hemorrhaging, abnor-
malities, etc.
e.g., tumors, hydrocephalus, hemorrhaging, abnor-
malities, etc.
A range of 8.25 cm is predefined.
* The conversion factor for an 8-week-old child, and a
scan range of 82.5 mm was used.
scan range of 82.5 mm was used.
HeadSeq
kV
80
Effective mAs
250
Rotation time
1.5 sec
Slice collimation
2.5 mm
Slice width
2.5 mm
Feed/Scan
5.0 mm
Kernel
C30s
CTDI
Vol
16.18 mGy
Effective dose
Male: 1.71 mSv*
Female: 1.87 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 196 Friday, April 8, 2005 9:55 AM