Siemens Software version syngo CT 2005C 用户手册

下载
页码 321
196
 
Children
HeadSeq_Baby
Indications:
Sequential mode for routine head studies for children, 
e.g., tumors, hydrocephalus, hemorrhaging, abnor-
malities, etc.
A range of 8.25 cm is predefined.
* The conversion factor for an 8-week-old child, and a 
scan range of 82.5 mm was used.
HeadSeq
kV
80
Effective mAs
250
Rotation time
1.5 sec
Slice collimation
2.5 mm
Slice width
2.5 mm
Feed/Scan
5.0 mm
Kernel
C30s
CTDI
Vol
16.18 mGy
Effective dose
Male: 1.71 mSv*
 
Female: 1.87 mSv*
Contrast medium IV injection
Start delay 
exam dependent
Flow rate 
dependent upon needle size/
Access site
Total amount 
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 196  Friday, April 8, 2005  9:55 AM