Siemens Software version syngo CT 2005C 用户手册

下载
页码 321
200
 
Children
InnerEar
Indications:
High Resolution Spiral mode for inner ear studies, e.g., 
malformations of the inner ear, inflammatory 
changes, pathologies of the mastoid process, tumor 
processes of the pyramids, post-traumatic changes, 
etc.
A range of 2.5 cm will be covered in 22 sec.
* The conversion factor for a 7-year-old child, and a 
scan range of 25 mm was used.
InnerEar
kV
80
Effective mAs
125
Rotation time
1.5 sec
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
2.0 mm
Pitch Factor
1.0
Increment
1.0 mm
Kernel
H80s
CTDI
Vol
8.09 mGy
Effective dose
Male: 0.04 mSv*
 
Female: 0.04 mSv*
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 200  Friday, April 8, 2005  9:55 AM