ThoraxHRSeq_Baby
Indications:
Sequence mode for High Resolution lung studies, e.g., 
interstitial changes of the lung parenchyma, etc.
A range of 20.1 cm is predefined.
* The conversion factor for an 8-week-old child, and a 
scan range of 201 mm was used.
ThorHRSeq
kV
80
Effective mAs
90
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
10.0 mm
Kernel
U90s
CTDI
Vol
0.50 mGy
Effective dose
Male: 0.88 mSv*
 
Female: 1.06 mSv*