Siemens Software version syngo CT 2005C 用户手册

下载
页码 321
234
 
Children
BodyAngio
Indications:
Spiral mode for abdominal CT Angio studies, e.g., vas-
cular abnormalities, aneurysms, etc.
A range of 10 cm will be covered in 19 sec. 
* The conversion factor for a 7-year-old child, and a 
scan range of 100 mm was used.
 BodyAngio
2
nd 
reconstr.
kV
80
Effective mAs
41
Rotation time
1.0 sec
Slice 
collimation
1.5 mm
Slice width
3.0 mm
2.0 mm
Feed/Rotation
6.0 mm
Pitch Factor
2.0
Increment
3.0 mm
1.5 mm
Kernel
B41s
B41s
CTDI
Vol
1.13 mGy
Effective dose
Male: 
0.73 mSv*
 
Female: 
0.84 mSv*
Contrast medium IV injection
Start delay 
exam dependent
Flow rate 
dependent upon needle size/
Access site
Total amount 
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 234  Friday, April 8, 2005  9:55 AM