syngo Osteo CT
 
287
Additional Important 
Information
• Fractured vertebrae are not suited for syngo Osteo 
CT evaluation since the more compact nature of 
these vertebrae result in bone mineral density value 
that is much higher than one would expect.
• When you examine a patient who is older than 80 
years, the patient examination data cannot be com-
pared with the reference data base.
• It is recommended to film directly from the Osteo 
card. Select images or series with Edit/Select all, and 
click “film” icon. You can also configure your film lay-
out, e.g., 3 x 3 as shown in Option/Configuration/
FilmingLayout.
Filming layout 
Note: it is not recommended to use filming layout set-
ting of 4 x 5 segments since the image text elements 
of the result image are overlapped and hard to read.
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 287  Friday, April 8, 2005  9:55 AM