Dose Information
Effective mAs 
In sequential scanning, the dose (Dseq) applied to the 
patient is the product of the tube current-time (mAs) 
and the CTDI
per mAs:
In spiral scanning, however, the applied dose (Dspiral) 
is influenced by the “classical“ mAs (mA x Rot Time) 
and in addition by the Pitch Factor. For example, if a 
Multislice CT scanner is used, the actual dose applied 
to the patient in spiral scanning will be decreased 
when the Pitch Factor is larger than 1, and increased 
when the Pitch Factor is smaller than 1. Therefore, the 
dose in spiral scanning has to be corrected by the Pitch 
Factor:
To make it easier for the users, the concept of the 
“effective“ mAs was introduced with the SOMATOM 
Multislice scanners.
The effective mAs takes into account the influence of 
pitch on both the image quality and dose:
To calculate the dose you simply have to multiply the 
CTDI
w
 per mAs with the effective mAs of the scan:
D
seq
 = D
CTDIw
 x mAs
D
spiral
 = (D
CTDIw
 x mA x Rot Time)/Pitch Factor
Effective mAs = mAs/Pitch Factor
D
spiral
 = D
CTDIw
 x effective mAs