Chapter 5
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Factory Preset: For each range, the following settings are selected:
Range State
(On),
Zero
Span
,
Res BW Mode
(Man),
Video BW
(Man) is set to 2 kHz, and
Sweep Time
Mode
(Auto) is set to 10 ms. The table below defines the other default
parameter settings for the measurement.
History:
Added with firmware revision A.03.00
Remote Command:
There is no equivalent remote command.
5.16.6.1
Range
Selects a range and updates the values on the other
Range Table
keys so that they reflect the
settings for the selected range. If
Range
is set to
On
it is used as part of the measurement;
when set to
Off
it is excluded. A range is made up of the next parameters.
Key Path:
Meas Setup
,
Range Table
Factory Preset: See
.
Range:
Off|On
Table 5-1
Range Table Default Settings
Range
Center
Freq (GHz)
Res BW
(kHz)
1
1.000
100.0
2
2.000
200.0
3
3.000
300.0
4
4.000
430.0
5
5.000
510.0
6
6.000
620.0
7
7.000
750.0
8
8.000
820.0
9
9.000
910.0
10
10.00
1000.0