National Instruments SCC-LP04 用户手册

下载
页码 18
SCC-LP Series Lowpass Filter Modules User Guide
14
ni.com
The horizontal axes of the first two plots are normalized to the SCC-LPXX 
cutoff frequency. When the input frequency (f) equals the cutoff frequency 
(fc), the normalized frequency has a value of 1.
The vertical axis of the third plot is normalized to the magnitude of the step 
input voltage. When the step-response output voltage equals the magnitude 
of the step-input voltage, the normalized step response is 1 V
OUT
/V
IN
.
Figure 6a shows that the SCC-LPXX provides 80 dB attenuation above 
ten times the cutoff frequency. Figure 6b shows variation in the group 
delay of the SCC-LPXX. Figure 6c shows the SCC-LPXX response to a 
step input. As shown, the peak voltage of the output is greater than the 
magnitude of the step input. If you expect step inputs, choose a gain setting 
and input range on the E Series DAQ device that allow for the effects of 
ringing. Otherwise the DAQ device input may be saturated, resulting in 
invalid data.