SCC-LP Series Lowpass Filter Modules User Guide
6
ni.com
Using the SCC-LPXX when Scaling Voltage Measurements
If you configured the SCC-LPXX using Measurement & Automation 
Explorer (MAX) and you are using NI-DAQ, the voltage reading you get 
from the E Series DAQ device accounts for the voltage scaling effect of the 
SCC-LPXX. Otherwise, you must scale the readings as follows:
1.
Read the SCC-LPXX channel on the E Series DAQ device 
V
ESERIES
(CHX).
2.
Calculate the SCC-LPXX voltage using this formula:
V
LP
 = 2V
ESERIES
where
V
LP
 is the SCC-LPXX input voltage.
V
ESERIES
 is the E Series DAQ device input voltage. 
Specifications
These ratings are typical at 25 °C unless otherwise stated.
Amplifier Characteristics
Number of input channels.......................2 DIFF
Input signal range
...................................±10 V
Output signal range
.................................±5 V
Gain ........................................................0.5
Input impedance .....................................10 G
Ω in parallel with 10 pF 
(powered on)
10 k
Ω 
(powered off or overloaded)
Gain error................................................Adjustable to 0% of reading
Offset-voltage error ................................350 
µV typ 
(referred to input, RTI)
1
1.5 mV max
2
1   
This specification is calculated relative to the input range of the module.
2   
Applicable at 25 °C.