BLENDEr
HELIUM-OxygEn
SPECIFICATIONS 
continued
dimensions: 
(without fittings)
 
Depth:
 
4.9 in
 
(12.5 cm)
 
Width:
 
2.3 in
 
(5.7 cm)
 
Height:
 
4.1 in
 
(10.4 cm)
Weight:
 
2.9 lbs  
(1.3 kg)
Shipping Weight: 
3.5 lbs  
(1.6 kg)
Operating Temperature Range:
59°F to 104°F (15°C to 40°C)
Transport / Storage Requirements
Temperature Range:
-10°F to 140°F (-23°C to 60°C)
Humidity:
Max 95% Noncondensing
F
I
O
2
 Accuracy:*
± 3% of full scale @ 50 psi (3.4 bar)
Pressure drop:
Low Flow: 
 2 psi (0.14 bar) at inlet pressures from 30-90 psi 
(2.1- 6.2 bar) and at 10 lpm flow rate at 60% F
I
O
2
.
High Flow: 
 3 psi (0.21 bar) at inlet pressures from 30-90 psi 
(2.1- 6.2 bar) and at 30 lpm flow rate at 60% F
I
O
2
.
The Helium-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Helium-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The alarmed Oxygen Monitor / Analyzer should comply with ISO 7767 to 
meet CE requirement.
dryness and Composition for inlet gases:
Heliox:
Medical grade 80/20 or 70/30 is required.
Oxygen:
Oxygen  supply  must  meet  all  requirements  of  USP 
Medical Oxygen.
dew Point:
(ONLY for CE 
 requirements)
Both  inlets  should  remain  10°F  (-12.2°C)  or  more 
below  the  lowest  temperature  to  which  the  air 
distribution  system  equipment  is  exposed.  At  a 
temperature  of  25°F  (-3.9°C)  and  a  pressure  of  
90 psi (6.33 kg/cm
2
) this equates to 2000 mg/m
3
.
* Accuracy of F
I
O
2
 
will be affected if bleed flow is not engaged at low flows. (At or below 3 lpm for Low Flow and 15 lpm for High Flow).
* When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow 
to maintain accuracy.
Specifications are subject to change without prior notice.