PRE-USE TESTING
WARNING
•  Read this User Manual before installing or operating the 
H
ELI
O
2
 Blender.
•  Confirm the concentration of heliox with an alarmed Oxygen Monitor /
Analyzer.
•  Confirm contents of heliox cylinder prior to use.
CAUTION
Inspect the H
ELI
O
2
 Blender for visual damage before use, DO NOT 
USE if damaged.
NOTE: The tests listed below should be performed prior to placing the 
H
ELI
O
2
 
Blender in service.
Pre-Use Testing consists of:
Alarm Test 
• Reverse Gas Flow Procedure
1. Secure the 
H
ELI
O
2 
Blender to a wall or pole bracket in an upright position.
2. Connect the heliox and oxygen supply lines to the appropriate inlet fittings 
on the bottom of the 
H
ELI
O
2
 Blender.
3. Attach a flowmeter, or other metering device to one of the outlet ports and 
verify 
F
I
O
range for accuracy with an alarmed Oxygen Monitor / Analyzer.
Primary Outlets Flow capacity:
• High Flow Blender (PM 5500 Model) 15 lpm to 120 lpm
• Low Flow Blender (PM 5400 Model) 3 lpm to 30 lpm
Auxiliary Outlet Use: 
The auxiliary flow outlet maintains the same flow capacity and 
F
I
O
2  
accuracy 
as the Primary Outlets with Bleed Valve not engaged. When the bleed flow is 
activated, some of the heliox mixture will vent to atmosphere to maintain 
F
I
O
 
concentration
 accuracy at the low flow settings.
High Flow Blender (PM 5500 Model) 15 lpm or less
• Low Flow Blender (PM 5400 Model) 3 lpm or less
4. Attach a supply line to the outlet port of the flowmeter.
ALARM TEST
1. Connect the 
H
ELI
O
2
 Blender to respective heliox and oxygen sources, 
pressurize the Blender and turn “ON” the flowmeter. 
2. Set Oxygen Concentration Dial to 50% 
F
I
O
2.
3. Disconnect or turn “OFF” the heliox supply line to the 
H
ELI
O
2
 
Blender. The 
Blender should alarm with a loud whistle noise. The whistle indicates the 
alarm is operating correctly.
4. Reconnect and activate the heliox supply line to the Blender, the alarm 
should stop whistling.
5. Disconnect or turn “OFF” the oxygen supply line to the 
H
ELI
O
2
 
Blender
The Blender should alarm with a loud whistle noise. The whistle indicates 
the alarm is operating correctly.
6. Reconnect and activate the oxygen supply line to the Blender, the alarm 
should stop whistling.
7. If alarm fails to function properly, DO NOT USE.