Chapter 2
Function Reference — AO_Calibrate
©
 National Instruments Corporation
2-25
NI-DAQ FRM for PC Compatibles
AO_Calibrate
Format
status = AO_Calibrate (deviceNumber, operation, EEPROMloc)
Purpose
Loads a set of calibration constants into the calibration DACs or copies a set of calibration 
constants from one of four EEPROM areas to EEPROM area 1. You can load an existing set of 
calibration constants into the calibration DACs from a storage area in the onboard EEPROM. 
You can copy EEPROM storage areas 2 through 5 (EEPROM area 5 contains the factory 
calibration constants) to storage area 1. NI-DAQ automatically loads the calibration constants 
stored in EEPROM area 1 the first time a function pertaining to the AT-AO-6/10 is called.
Note
Use the calibration utility provided with the AT-AO-6/10 to perform a calibration 
procedure. Refer to the calibration chapter in the 
AT-AO-6/10 User Manual for 
more information regarding the calibration procedure.
Parameters
Input 
Parameter Discussion
operation determines the operation to be performed.
1:
Load calibration constants from EEPROMloc.
2:
Copy calibration constants from EEPROMloc to EEPROM user calibration 
area 1.
EEPROMloc selects the storage location in the onboard EEPROM to be used. You can use 
different sets of calibration constants to compensate for configuration or environmental 
changes.
1:
User calibration area 1.
2:
User calibration area 2.
3:
User calibration area 3.
4:
User calibration area 4.
5:
Factory calibration area.
Name
Type
Description
deviceNumber
i16
assigned by configuration utility
operation
i16
operation to be performed
EEPROMloc
i16
storage location in the onboard EEPROM