Tektronix AWG710B 用户手册

下载
页码 580
The Setup Menu Screen
AWG710&AWG710B Arbitrary Waveform Generator User Manual
3-53
Starting and Stopping
Output
When you load or create a waveform in the waveform memory, output does not 
start until you push the RUN button on the front–panel. The RUN LED is on and 
the instrument starts sweeping the waveform data in the waveform memory.
When the Waveform Generator is set to the Trigger mode, the Waveform 
Generator waits for a trigger event to be generated by pushing the FORCE 
TRIGGER
 button or by external trigger event signal. Refer to The Run Mode 
Menu
 on page 3-44.
The current run state of the instrument is displayed in the status area at the upper 
part of the screen as shown in Figure 3-5 page 3-44. Refer to Table 3-7 for state 
messages.
If waveforms are not present, the Running 
or 
Waiting message will be changed to 
the Stopped message. There will be no output when you push the RUN button and 
the Stopped message will continue to be displayed.
The RUN LED is on when the run state is Running 
or 
Waiting
.
The line circuit from internal generator module to the output connector must be 
closed to output waveform from the front–panel output connector. 
Turning Channel Output
On and Off
Push the CH1 OUT button to connect or disconnect the instrument output to the 
CH 1 connector. When you push the CH1 button, the CH1 LED goes on and a 
waveform is output from the CH1 connector if the instrument is in the Running 
state. When you push the CH1 button again, the signal output is disconnected and 
the waveform output is stopped, even if the instrument is in the Running state. If 
there is no waveform loaded into a channel, you cannot turn that channel output on 
or off.
The CH1 LED automatically turns off when the waveform data in that channel 
becomes invalid. For example, you attempt to load an incorrect file, and the 
instrument deletes the current waveform from memory.
Table 3-7: Instrument run state and state messages
State messages
Descriptions
Stopped
The output operation is currently stopped.
Waiting
The instrument is waiting for a trigger or an event signal.
Running
The instrument is outputting waveform(s).