90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 35
• 
Change with the 
↓↓
 -key to the next line to put in the second value.
• 
Repeat these steps until all raw/setpoint values of the table are part of the
corresponding menus.
• 
If you have more than 10 values you can go to further menu pages with the F5 key
(More...). There are maximum 30 lines available. The current menu page number is
always displayed in the headline. If you have less than 30 values you have to put in
"0" in the line following the last real value of the table!
Notes:
♦ 
You can change between the menus "Linearization Raw Values" and "Linearization
Setpoint Values" with the F2 key!
♦ 
The raw value "X2" must belong to the setpoint value "Y2" ! Otherwise, the calculation
of the linearization may be wrong !
Measure
Y1...10
Channel
Back...
More...
TAG
F1
F2
F3
F4
F5
3163.860 ppm
-- Linearization Raw Values (1/3) --
X1:
0.000  ppm
X3:
387.709  ppm
X4:
517.464  ppm
X5:
645.199  ppm
X6:
757.313  ppm
X7:
1113.910  ppm
X9:
3163.860  ppm
X10:
0.000  ppm
X2:
266.291  ppm
X8:
2263.390  ppm
Measure
X1...10
Channel
Back...
More...
TAG
F1
F2
F3
F4
F5
3163.860 ppm
-- Linearization Setpoint Values (1/3) --
0.000  ppm
Y3:
Y4:
428.610  ppm
536.760  ppm
Y6:
Y7:
955.395  ppm
3163.860  ppm
Y10:
Y2:
217.455  ppm
2105.560  ppm
5.1 Analyzer Module Setup
5.1.5 Linearization