Emerson I-0021 用户手册

下载
页码 238
5 - 78
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Function Codes 1 - 599
Available for:
♦ 
The following functions occur only if the signal at the input changes from low to high level:
ID
General functions
1
Set all channels into STANDBY state (Stop procedures)
2 - 99
Reserved
ID
CH1
 ID
CH2
 ID
CH3
 ID
CH4
 ID
CH5
Channel dependent
functions
100 
200 
300 
400 
500 
Start Zero calibration
101 
201 
301 
401 
501 
Start Span calibration
102 
202 
302 
402 
502 
Start Zero + Span calibration
103 
203 
303 
403 
503 
Open sample gas valve
104 
204 
304 
404 
504 
Open zero gas valve
105 
205 
305 
405 
505 
Open purge gas valve
106 
206 
306 
406 
506 
Open test gas valve
107 
207 
307 
407 
507 
Open linearizer gas valve
108 
208 
308 
408 
508 
Open span gas valve (of current range)
109 
209 
309 
409 
509 
Open span gas valve of range-1
110 
210 
310 
410 
510 
Open span gas valve of range-2
111 
211 
311 
411 
511 
Open span gas valve of range-3
112 
212 
312 
412 
512 
Open span gas valve of range-4
113 
213 
313 
413 
513 
Close all valves (STANDBY)
114 
214 
314 
414 
514 
Set range-1
115 
215 
315 
415 
515 
Set range-2
116 
216 
316 
416 
516 
Set range-3
117 
217 
317 
417 
517 
Set range-4
118 
218 
318 
418 
518 
Reserved
...
...
...
...
...
...
149
249
349
449
549
Reserved
The following functions are active if the signal at the input is set to high level:
ID
CH1
 ID
CH2
 ID
CH3
 ID
CH4
 ID
CH5
Channel dependent
functions
150 
250 
350 
450 
550 
AK key-switch: Remote
151 
251 
351 
451 
551 
Namur signal: Function Control
152 
252 
352 
452 
552 
Reserved
...
...
...
...
...
...
199
299
399
499
599
Reserved