National Instruments NI PXI/PCI-5114 用户手册
NI PXI/PCI-5114 Specifications
14
ni.com
Calibration
Power
Specification
Value
Comments
Self-Calibration
Self-calibration is done on software command.
The calibration corrects for gain, offset,
compensated 1 M
The calibration corrects for gain, offset,
compensated 1 M
Ω attenuator, triggering, and
timing adjustment errors for all input ranges.
—
External Calibration
(Factory Calibration)
(Factory Calibration)
The external calibration calibrates the VCXO, gain,
and the voltage reference. Appropriate constants are
stored in nonvolatile memory.
and the voltage reference. Appropriate constants are
stored in nonvolatile memory.
—
Interval for External
Calibration
Calibration
2 years
—
Warm-Up Time
15 minutes
—
Specification
Typical Value
Comments
+3.3 VDC
NI PXI-5114
NI PCI-5114
—
840 mA
1.6 A
+5 VDC
1.1 A
1.7 A
+12 VDC
250 mA
45 mA
–12 VDC
170 mA
—
Total Power
13.32 W
14.32 W