NI 5412 Specifications
18
ni.com
Calibration
Power
Table 10.  
Specification
Value
Comments
Self-Calibration An onboard, 24-bit ADC and precision voltage reference are 
used to calibrate the DC gain and offset. The self-calibration 
is initiated by the user through the software and takes 
approximately 75 seconds to complete.
External 
Calibration
The External Calibration calibrates the VCXO, voltage 
reference, DC gain, and offset. Appropriate constants are 
stored in nonvolatile memory.
Calibration 
Interval
Specifications valid within 2 years of External Calibration.
Warm-up Time
15 minutes
Table 11.  
Specification
Normal Operation
Overload Operation
Comments
Total Power
22 W
26 W
Typical.
Overload 
operation occurs 
when CH 0 is 
shorted to 
ground.