NI 5412 Specifications
NI PCI-5412 Environment
Note
To ensure that the NI PCI-5412 cools effectively, follow the guidelines in the
Maintain Forced-Air Cooling Note to Users included in the NI 5412 kit. Also, to maximize
airflow and extend the life of the device, leave any adjacent PCI slots empty. The
NI PCI-5412 is intended for indoor use only.
airflow and extend the life of the device, leave any adjacent PCI slots empty. The
NI PCI-5412 is intended for indoor use only.
Table 14.
Specifications
Value
Comments
Operating
Temperature
Temperature
0 ºC to +45 ºC. Meets IEC-60068-2-1 and IEC-60068-2-2.
—
Storage
Temperature
Temperature
–25 ºC to +85 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
IEC-60068-2-2.
—
Operating
Relative
Humidity
Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
—
Storage
Relative
Humidity
Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
—
Storage
Shock
Shock
50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
profile developed in accordance with MIL-PRF-28800F.
—
Storage
Vibration
Vibration
5 Hz to 500 Hz, 2.46 g
rms
. Meets IEC-60068-2-64. Test
profile exceeds requirements of MIL-PRF-28800F, Class B.
—
Altitude
2,000 m maximum (at 25 °C ambient temperature)
—
Pollution
Degree
Degree
2
—