Clarity 87392-11 用户手册

下载
页码 53
Clarity User Guide,
 
March 1, 2011 - Rev. 1.06 
                    
           Page 33
On the Calibration screen under Step 4 (See Fig. 4.1), there are six parameters that describe 
the lens that will be analyzed: 
 
1. Diameter 
2. 
Index of Refraction 
3. Power 
4. Front 
Curve 
5. Back 
Curve 
6. Center 
Thickness 
 
These parameters are for both the plus and the minus lenses. 
 
      NOTE 
 
If these lenses are misplaced or compromised in any way, please call the factory and 
order a new calibration lens.  Once the new lens arrives, check these parameters and 
make any necessary changes. 
 
To perform this calibration, select the Calibrate button.  You will be prompted to remove the 
lens.  Make sure there is no lens in the chamber and select OK.  After 25 seconds, you will be 
prompted to insert the calibration lens.  Put the -6.00 lens in the chamber and select OK.  The 
Clarity will analyze the lens and will instruct you to place the +6.00 lens in the chamber.  
Select OK
 
After both lenses are analyzed, the plus and minus slope adjustments will be displayed.  Select 
YES to accept these changes.  The calibration wizard will appear telling you that the power 
calibration is complete.  Select OK.  To exit back to the Job screen, select OK at the bottom 
of the screen and then on the next screen select Next. 
 
      NOTE 
 
On this screen, there is also a Power Offset box.  This can be used if you see a consistent 
power error.  For example,  powers are showing 0.25 weak across the board.  Enter a 
+.25 and all analyzed powers will have 0.25 added to the final reading.  This functions 
like the manual eyepiece adjustment for power on a conventional lensometer. 
 
This screen includes two parameters labeled Minus Prism Bias and Plus Prism Bias.  
They are constants and should not be changed unless directed by the factory. 
 
 
To verify power calibration, there is an internal job (Job Number 999999999) which will 
allow you to use the two calibration lenses (+6 and -6).  From the main Job screen enter the 
Job Number 999999999 and follow the directions.  Each lens will show the measured power 
in the upper right hand corner of the screen.  If they are within acceptable limits, you can 
move on.  If they are out of tolerance, recalibrate.