Clarity 87392-11 用户手册

下载
页码 53
Clarity User Guide,
 
March 1, 2011 - Rev. 1.06 
                    
           Page 41
 
Chapter 6 
– Maintenance
 
 
 
      NOTE 
 
Always unplug unit before performing any maintenance. 
 
 
Cleaning the Glass Plate:   
 
Lift the glass plate out (See Fig. 1.5) and clean it using a microfiber cloth as used in cleaning 
A/R coated lenses.  You can also clean the glass plate with glass cleaner or warm soapy water 
and a clean water rinse.  Make sure that the plate is completely dry before replacing into the 
chamber.  Replace the glass plate and resume using the Clarity. 
 
Cleaning the LCD:   
 
Remove the glass plate (See Fig. 1.5) and reach down into the opening that the glass plate 
covered.  Using a microfiber cloth, clean the LCD gently and thoroughly.   
 
Cleaning the Touch Screen:   
 
The touch screen can be cleaned using a microfiber cloth.  Wipe the screen gently until 
fingerprints and smudges are eliminated. 
 
Replacement of the Main Fuse Located on the Rear Bulkhead (Fig. 1.3): 
 
Using a flat head screw driver, insert it into the slot on the cover and push in while rotating to 
remove the cover/holder.  Remove the bad fuse and replace it with the proper 250v 5amp 5x20 
mm slow blow fuse. Re-install the fuse holder using the flat head screwdriver to push in and 
rotate to lock in place. 
 
 
Maintenance Tips 
 
•  Do not attempt to lubricate any part of the mechanism except under supervision of 
factory service personnel. 
•  Do not interfere with the motion of the lens grippers or the blocking mechanism. 
•  Operate the equipment in an environment free from dust or debris. 
•  Use the burn-in function to warm the machine if its temperature is estimated to be 5
or 9
F below the final stable operating temperature. 
•  Do not ship the equipment unless you have re-installed the factory shipping restraints 
and foam packing-box inserts.