Clarity 87392-11 用户手册

下载
页码 53
Clarity User Guide,
 
March 1, 2011 - Rev. 1.06 
                    
           Page 44
 
 
•  Manual alignment of lenticular and occupational lenses is recommended.  
Occupationals could be aligned upside down in automatic mode.  
 
•  In the case of extreme prism >6 diopters, the prism accuracy is good, but the lens may 
be tilted to such a degree that cylinder accuracy decreases and spheres may show some 
cylinder.  This can be eliminated by holding the lens so that the front curve is parallel 
to the glass plate when you insert it into the chamber and then select START and let 
the clamps close on the lens in that position.  It will then process normally.