National Instruments NI 5422 用户手册

下载
页码 36
NI 5422 Specifications
24
ni.com
Start Trigger
DDC CLK IN
Clocking 
Format
DDC Data Output signals change on the rising edge of 
DDC CLK IN.
Frequency 
Range
10 Hz to 200 MHz
Input Duty 
Cycle 
Tolerance
40% to 60%
Table 8.  
Specification
Value
Comments
Sources
1. PFI<0..1> (SMB front panel connectors)
2. PFI<2..3> (DIGITAL DATA & CONTROL front panel 
connector)
3. PXI_Trig<0..7> (backplane connector)
4. PXI Star trigger (backplane connector)
5. Software (use function call)
6. Immediate (does not wait for a trigger). Default.
Modes
1. Single
2. Continuous
3. Stepped
4. Burst
Edge Detection
Rising 
Minimum 
Pulse Width 
25 ns. Refer to t
s1
 at NI Signal Generators Help»Devices»
NI 5422»NI PXI-5422»Triggering»Trigger Timing.
Table 7.   (Continued)
Specification
Value
Comments