National Instruments NI 5422 用户手册

下载
页码 36
© National Instruments Corporation
5
NI 5422 Specifications
Output Characteristics (Continued)
Output Enable
Software-selectable. When the Output Path is disabled, the 
CH 0 Output is terminated to ground with a 1 W resistor 
equal to the selected output impedance.
Maximum 
Output 
Overload
The CH 0 output can be connected to a 50 
Ω, ±12 V 
(±8 V for the Direct Path) source without sustaining any 
damage. No damage occurs if the CH 0 output is shorted to 
ground indefinitely.
Waveform 
Summing
The CH 0 output supports waveform summing among 
similar paths—specifically, the outputs of multiple NI 5422 
signal generators can be connected directly together.
Frequency and Transient Response
Analog 
Filter
Software-selectable 7-pole elliptical filter for image 
suppression.
Available on 
Low-Gain 
Amplifier and 
High-Gain 
Amplifier Paths.
Pulse 
Response
Path
Values are 
typical. Analog 
Filter disabled. 
Measured with a 
1 m RG-223 
cable.
Direct
Low-Gain 
Amplifier
High-Gain 
Amplifier
Rise/Fall Time
1.0 ns
2.1 ns
4.8 ns
Aberration
16%
6%
8%
Table 1.   (Continued)
Specification
Value
Comments