A&D GP-100K 用户手册

下载
页码 98
 
 29
 
 
 
5  Confirm that there is nothing on the pan and press 
the PRINT key. The balance measures the zero 
point. Do not allow vibration or drafts to affect the 
balance. 
  The balance displays the calibration weight value. 
6  Place the displayed calibration weight on the pan 
and press the PRINT key. The balance measures 
the calibration weight. Do not allow vibration or 
drafts to affect the balance. 
7  The balance displays  end . Remove the weight 
from the pan. 
8 If the “GLP output (info)” parameter, of the 
function table, is set to “1” or “2”, the balance 
displays  glp  and outputs “Calibration Report” 
using the RS-232C interface or stores the data in 
memory. For details on the calibration report format, 
refer to “10-2 GLP Report”. 
9 The balance will automatically return to the 
weighing mode. 
10  Place the calibration weight on the pan and confirm 
that the value displayed is within 
±2 digits of the 
specified value. If it is not within the range, check 
the ambient conditions such as breeze and 
vibration, also check the weighing pan. Then, 
repeat steps 1 to 10.