The FMG210/220/230 Series are small electromagnetic 
low-flow  flowmeters  with  chemically-resistant  plastic  wet-
ted parts and a choice of Hastelloy or platinum electrodes 
suitable  for  use  with  a  variety  of  chemicals.    Capable  of 
measuring  pulsating  flows  from  diaphragm-type  metering 
pumps, they are designed primarily for electrically-conduc-
tive chemical injection applications. The 1/4”, 3/8”, and 
1/2” sizes monitor maximum flows of 1, 3, and 8 gallons 
per minute (or 4, 11, and 30 L/min), respectively. Barb or 
NPT fittings are available.
The FMG210/220/230 Series are 12-24 Vdc powered and 
come with standard pulse and analog outputs.  All electron-
ics are self-contained in a single compact chemically-resis-
tant housing, well suited for OEM applications.
These meters are also appropriate for remote monitoring of 
chemical volume or rate using a computer or a PLC.
GENERAL INFORMATION
FEATURES
SPECIFICATIONS
6 ft. control cable
Materials  Body
 
Electrodes
 
O-Ring
 
Fittings
Power
Accuracy
Max Fluid Temperature
Maximum Pressure
Minimum Conductivity
Outputs
ABS plastic
Platinum or Hastelloy
Viton™ (EPDM optional)
PVDF plastic
12-24 Vdc, 180 mA
+/-1% of reading, plus .005 gpm (0.02 lpm)
185˚ F (85˚ C)
150 psi (@ 75˚ F)
20 microSiemens
• Frequency opto isolated 1000 pulse/liter (3785 p/g)
• 4-20 mA opto isolated passive
Strain relief
Rugged ABS plastic body
Male NPT or barb pipe fittings,
PVDF plastic
FLOw RANGE
  SIZE 
LOw FLOw CUTOFF 
MINIMUM FLOw for spec accuracy 
MAXIMUM FLOw
-025
-038
-050
 
L/MIN 
GAL/MIN 
GAL/HR 
L/MIN 
GAL/MIN 
GAL/HR 
L/MIN 
GAL/MIN 
GAL/HR
 
.04 
.01 
.60 
.26 
.07 
4.2 
3.8 
60
 
.11 
.03 
1.80 
.76 
0.2 
12 
11 
180
 
.26 
.07 
4.20 
1.9 
0.5 
30 
30 
480