Foreword
Notes on Use
•  This instrument should be used under the following operat-
ing conditions.
  Temperature of 0 to 40°C (32 to 104°F), relative humidity 
of 30 to 85%, with no condensation.
  Atmospheric pressure of 700 to 1060 hPa (altitude of 
400 to 3000m).
•  This is a precision instrument.  To avoid the possibility of it 
being damaged, the instrument should not be dropped nor 
should heavy objects be placed on top of it.
•  Do not expose this instrument to rain and water.
•  This instrument has no alarm function. Do not use this in-
strument if an alarm function is  necessary.
•  This instrument is designed for use on adults. When using 
the instrument on infants or babies, please note that the de-
signed measurement accuracy cannot be guaranteed.
•  Do not use the instrument on a point where there is exces-
sive vibration. In addition, to avoid excessive impact on the 
instrument, it should be handled gently. Failure to observe 
this may cause breakdown.
•  Accurate measurement data may not be obtained in the fol-
lowing cases. 
  When body movement is excessive
  When the probe is not attached properly
  When blood circulation at the measuring point is poor (due 
to pressure on arm/finger, peripheral circulatory insuffi-
ciency)
  When ambient light (e.g. panel light, fluorescent light, in-
frared heat lamp, direct sunlight) is too strong
  When influenced electromagnetically by other electronic 
devices (e.g. near electrical appliances like TV, medial 
device)
  When a mobile telephone is used during measurement