National Instruments 4-Channel 用户手册
NI USB-9215 Series User Guide and Specifications
20
ni.com
Safety
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
NI USB-9215 with Screw Terminal Safety Voltages
Connect only voltages that are within these limits.
Channel-to-COM ....................................±30 V max
Isolation
Channel-to-channel..........................No isolation between channels
Channel-to-earth ground
Withstand .................................2,300 V
rms
, 5 seconds max
Continuous ...............................250 V
rms
,
Measurement Category II
Measurement Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category refers
to local-level electrical distribution, such as that provided by a standard
wall outlet (for example, 115 V for U.S. or 230 V for Europe). Do not use
this module with Measurement Category III or IV voltages.
directly connected to the electrical distribution system. This category refers
to local-level electrical distribution, such as that provided by a standard
wall outlet (for example, 115 V for U.S. or 230 V for Europe). Do not use
this module with Measurement Category III or IV voltages.
NI USB-9215A with BNC Safety Voltages
Connect only voltages that are within these limits.
AI+-to-AI–..............................................±30 V max
Isolation
Channel-to-channel..........................No isolation between channels
Channel-to-earth ground
Withstand .................................1,500 V
rms
, 5 seconds max
Continuous ...............................60 VDC,
Measurement Category I
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics. Do not
use this module with Measurement Category II, III, or IV voltages.
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics. Do not
use this module with Measurement Category II, III, or IV voltages.