National Instruments SCXI -1125 用户手册

下载
页码 101
Appendix 
A  Specifications
A-8
ni.com
Maximum Working Voltage
Maximum voltage rating refers to the signal voltage plus the common 
mode voltage (Signal + common mode). Voltage of each input shall remain 
within 300 V of ground.
Channel-to-earth .....................................300 V, Measurement Category II
Channel-to-channel.................................300 V, Measurement Category II
Cautions 
The SCXI-1125 is rated for Measurement Category II and is intended to carry 
signal voltages no greater than 300 V. Do not use the SCXI-1125 for connection to signals 
or for measurements within Categories III or IV.
 
 
When hazardous voltages (>42.4 V
pk
/60 VDC) are present on any channel, all channels are 
considered hazardous. Ensure that external wiring or any circuits connected to the device 
are properly insulated from human contact.
Environmental
Operating temperature ............................0 to 50 
°C
Storage temperature ................................–20 to 70 
°C
Humidity .................................................10 to 90% RH, noncondensing
Maximum altitude...................................2,000 meters
Pollution Degree (indoor use only) ........2
Table A-2.  Terminal Block Maximum Voltages
Module
Signal Range
Maximum Voltage and Category [Insulation]
SCXI-1313A
150 V
150 V, Measurement Category II
 
[channel-to-channel, channel-to-earth]
SCXI-1327
300 V
300 V, Measurement Category II
 
[bank-to-bank, bank-to-earth]
TBX-1316
600 V
600 V, Measurement Category II
 
[Basic; channel-to-channel]
1000 V
1000 V, Measurement Category I
 
[Basic; channel-to-channel]